Logo - CWRDM National Stable Isotope Facility for Research in Basic and Applied Sciences Logo - Department of Science and Technology, Government of India
 
  IRMS | GC-MS | IC | ULLSC | HPLC

  Isotope Ratio Mass Spectrometer
innerbg2 innertopbg3
 
   
Instrument Specification  

Model:


FinniganTM DELTAplus XP Continuous Flow Isotope Ratio Mass spectrometer

Peripherals:

Gas Bench II with GC PAL auto sampler, Flash Elemental Analyzer and High Temperature Elemental Analyzer both coupled with Con Flow III
Isotope Ratio Mass Spectrometer

Application Area
:

Determination of environmental stable isotope ratios of D/H, 18O/16O, 13C/12C, 15N/14N and 34S/32S.
Recommended values and values obtained in this laboratory for the primary and  secondary standards, using FinniganTM DELTAplus XP Continuous Flow IRMS are given below:



Name of
Standard
Recommended Value Analyses Value (National Facility)
for delta D
(per mil)
for delta 18 O
(per mil)
for delta D
(per mil)
for delta 18 O
(per mil)
VSMOW 0.00 0.00 0.01 0.006
GISP -189.73 -24.78 -188.85 -24.98
SLAP -428.00 -55.49 -429.13 -55.49
NARM -35.86 -4.68 -35.80 -4.66

The data reproducability for Hydrogen is 0.25‰ and that for Oxygen is 1‰